• GREISINGER Atmospheric oxygen transducer OXY 3690 MP

GREISINGER Atmospheric oxygen transducer OXY 3690 MP

Measuring ranges
Oxygen concentration: 0.0 ... 100.0 % O2 (gaseous)
OXY 3690 MP: recommended range 0.2 ... 35.0 vol.% O2 (reduced precision outside)
OXY 3690 MP-LO: also suitable for values ≤0.2 vol.% O2
Temperature: -20.0 ... 50.0 °C
Accuracy device (at nominal temperature 25 °C):
Oxygen: ±0.1 % ±1 digit
Temperature: ±0.1 °C ±1 digit
Output signal (O2 only): 4 ... 20 mA (2-wire - standard), 0 ... 10 V (3-wire - option)
Model : GREISINGER-OXY3690MP

Oxygen transmitter, for air oxygen measurements, incl. sensor

THE DEVICE IS ONLY INTENDED FOR CONTROL. IT IS NOT A REPLACEMENT FOR A MONITORING DEVICE SUBJECT TO AUTHORISATION!
Measuring ranges
Oxygen concentration: 0.0 ... 100.0 % O2 (gaseous)
OXY 3690 MP: recommended range 0.2 ... 35.0 vol.% O2 (reduced precision outside)
OXY 3690 MP-LO: also suitable for values ≤0.2 vol.% O2
Temperature: -20.0 ... 50.0 °C
Accuracy device (at nominal temperature 25 °C):
Oxygen: ±0.1 % ±1 digit
Temperature: ±0.1 °C ±1 digit
Output signal (O2 only): 4 ... 20 mA (2-wire - standard), 0 ... 10 V (3-wire - option)
Electric isolation: input electrically isolated
Auxiliary energy: 12 ... 30 V DC (at output 4 ... 20 mA) / 18 ... 30 V DC (at output 0 ... 10 V - option)
Perm. impedance (at 4 ... 20 mA):RA [Ω] ≤ (Uv [V] - 12 V) / 0.02 A
Permissible load (at 0 ... 10 Volt):RL >3000 Ω
Working condition: 0 ... +50 °C, 0 ... 95 % RH (non-condensing)
Storage temperature: -20 ... +70 °C
Reverse voltage protection:50 V permanently
Display: approx. 10 mm high, 4-digit LCD-display
Housing: ABS (IP65 - with the exception of sensor plug)
Dimensions: 82 x 80 x 55 mm (without elbow-type plug and sensor plug)
Electric connection: elbow-type plug acc. to EN 175301-803/A (IP65), max. wire cross section: 1.5 mm², wire diameter from 4.5 ... 7 mm
Sensor connection: 5-pin jack connector, screwable
Calibration: 1-point calibration in atmospheric air
Air pressure compensation: 500 ... 2000 hPa abs., manually input

Oxygen sensor:
Sensor type: depending on the version, see above
Measuring ranges: 0.0 ... 100.0 % O2
Response time T90: <10 s, depending on temperature
Warranty: 12 months (assuming appropriate usage according to the
manual)
Application area: suitable for air and pure oxygen, protective gases
Temperature compensation: integrated in sensor housing
Connection cable: approx. 1.3 m, with 5-pin plug, screwable
Operating pressure: 500 ... 2000 hPa (static).
For air and gas-stream use the option GOO.../MU.
Working condition: 0 ... +45 °C, 0 ... +95 % RH (non-condensing)
Storage temperature: -15 ... +60 °C