Flow calibration

CALIBRATION LABORATORY

Flow Standard Calibration Laboratory

ILAC Recognition| TAF (ISO/IEC 17025) Accredited Laboratory No. 3032

Flow Rate Range 2.54 m³/h ... 30.0 m³/h
Flow Velocity Range 0.2 m/s ... 3 m/s
Smallest Uncertainty Flow Rate 0.9%|Flow Velocity 0.08 m/s

Features

  • Flow calibration ranges:
    • Flow rate: 2.54 m³/h ... 30.0 m³/h (42.3 L/min ... 500 L/min); smallest uncertainty: 0.9%.
    • Flow velocity: 0.2 m/s ... 3 m/s; smallest uncertainty: 0.08 m/s.
  • Low calibration uncertainty and excellent repeatability, suitable for various types of flowmeters and flow transmitters.

Flow Calibration Conditions

  • Applicable types include insertion-type hot-wire, ultrasonic, electromagnetic, Coriolis, differential-pressure, turbine, and vortex flowmeters.
  • The calibration medium is tap water.
  • Available measuring-pipe sizes are DN40 (1.5 inches) and DN80 (3 inches).
  • Insertion-type flowmeters must meet the following conditions:
    1. The probe outer diameter must be φ12 mm or smaller.
    2. The measuring-point insertion depth for a DN40 pipe is 21.3 mm.
    3. The measuring-point insertion depth for a DN80 pipe is 40.9 mm.

Calibration System Architecture Diagram

Flow Standard Calibration Laboratory system architecture

Accredited Capabilities

Flow Calibration LaboratoryWorking StandardCheck Standard
Calibration InstrumentYOKOGAWA Magnetic FlowmeterTRICOR Coriolis Flowmeter
Scope of AccreditationFlow rate: 2.54 m³/h ... 30.0 m³/h (42.3 L/min ... 500 L/min); flow velocity: 0.2 m/s ... 3 m/s
Smallest UncertaintyFlow rate: 0.9%; flow velocity: 0.08 m/s