Temperature calibration

CALIBRATION LABORATORY

Temperature Standard Calibration Laboratory

ILAC Recognition| TAF (ISO/IEC 17025) Accredited Laboratory No. 3032

Calibration Item Resistance Thermometer
Scope of Accreditation -40°C ... +300°C
Smallest Uncertainty 0.029°C ... 0.085°C

Features

  • Resistance thermometer calibration range: -40°C ... +300°C; smallest uncertainty: 0.029°C ... 0.085°C.
  • A precision and stable Fluke calibration system is used together with a comprehensive verification mechanism.
  • Low calibration uncertainty and excellent repeatability, suitable for various resistance thermometers and thermocouple thermometers.

Temperature Calibration Conditions

  • The calibration item must include a display, meter, or indicator.
  • Probe-type instruments only, with a thread size of 1/2 inch or smaller. Flange-type instruments are not accepted.
  • The probe length must be at least 10 cm, excluding the threaded section.
  • The outer diameter of the measuring probe must be φ12 mm or smaller.
  • The outer diameter of the instrument body must be φ34 mm or smaller.
  • A water bath is used for calibration below 20°C. The temperature-sensing element must not be exposed, and the protective sheath must prevent water ingress.

Calibration System Architecture Diagram

Temperature Standard Calibration Laboratory system architecture

Accredited Capabilities

Temperature Calibration LaboratoryResistance Thermometer
Working StandardChino Standard Platinum Resistance Thermometer
Scope of Accreditation-40°C ... +300°C
Smallest Uncertainty0.029°C ... 0.085°C