2012/07 Machinery Monthly - Development of an Automated Quality Inspection System for Humidity Sensor

Development of an Automated Quality Inspection System for Temperature and Humidity Sensors


Chia-Te Chou1 Ming-Chung Tsai1 Jen-Cheng Chu2


1Graduate Institute of Automation and Control, National Taiwan University of Science and Technology
2Yuden Control Technology Co., Ltd.


Abstract:This paper proposes an automated quality inspection system for temperature and humidity sensors. The system includes an automatically controlled humidity sensing chamber, a temperature and humidity sensor under test, a dew point hygrometer (reference standard), a USB data acquisition module, and a PC-based human–machine interface control system. The temperature and humidity sensing chamber is based on the dry–wet dual-pressure method. By adjusting the pressure value of the moisture valve through voltage output from the USB data acquisition module, a controllable humidity chamber is constructed to achieve the stable chamber humidity required for testing. Through program-based automatic calculation and adjustment of the moisture pressure valve, once the set humidity value reaches steady state, the system immediately measures both the device under test and the reference standard and compares the error values. Finally, the inspection results are exported to an EXCEL report and quality characteristic charts are generated to provide reference information for quality control personnel. Experimental verification shows that the automated inspection technology can effectively be introduced into the quality control process of temperature and humidity sensors, improving product reliability while reducing manual labor and inspection time.

Keywords: relative humidity, dew point, temperature and humidity sensors, quality inspection, automated data acquisition.

 

1. Introduction

Temperature and humidity are important physical parameters in environmental conditions, significantly affecting human living quality and industrial product processing quality. Therefore, temperature and humidity measurement and monitoring have become increasingly common across industries. In recent years, research related to temperature and humidity sensor development and anti-condensation technology has gradually increased [1,2]. Modern electronics manufacturing facilities require stricter control of temperature and humidity, generally requiring both temperature and relative humidity to be maintained within specified ranges. In general factories, moisture is often introduced into the space through air conditioning systems. Therefore, the humidification and dehumidification performance of the air conditioning system is determined by the moisture delivered into the space. Especially under low humidity conditions, accurate dew point measurement becomes particularly important for temperature and humidity control. When dew point temperature fluctuates, relative humidity will naturally fluctuate as well, and temperature–humidity control is adjusted according to feedback values within the lower and upper limits.

In addition, within compressed air system circuits, pressure dew point measurement in compressed air or gas distribution networks has become increasingly important. Compressor manufacturers have gradually realized that excessive moisture in pneumatic pipelines may cause many negative effects, such as pipeline rust and corrosion, shortened lifespan of pneumatic components, actuator failures, contamination of the entire compressed air system, and damage to system components. These issues may lead to unplanned production interruptions and increased additional production costs [3]. If, due to maintenance factors, the performance of dryers used to remove moisture from gases cannot operate as expected, temperature–humidity or dew point measurements can serve as an auxiliary monitoring system to evaluate the dehumidification performance of dryers. Alarms can be triggered when measured values exceed preset limits. Therefore, the advantages of temperature–humidity or dew point measurements in compressed air systems include: (1) continuous monitoring of pressure dew point to detect and resolve problems in compressed air drying processes in time; (2) extending the service life of compressed air systems and their components; (3) improving maintenance efficiency of compressed air systems; and (4) reducing system operation failures to ensure stable product quality [3].

Environmental control usually requires regulation of both gas temperature and humidity. In reality, air contains water vapor and forms moist air. Common humidity measurement methods include chilled mirror technology, capacitive polymer sensors, and capacitive metal oxide sensors. Their measurement principles are briefly described as follows:

(1) Chilled mirror technology: When air flows across a mirror surface, the mirror gradually cools until condensation forms on the surface. A photosensor detects whether condensation has occurred. When the intensity of reflected light reaches a certain level, it indicates the presence of water droplets on the mirror surface, and the temperature measured at that moment is defined as the dew point temperature. However, the optical sensor used in this method must be highly sensitive. If contaminants such as dirt, oil, or dust adhere to the mirror surface, measurement results may be affected. Nevertheless, this method provides high measurement accuracy, though at higher cost, and is suitable for high-precision measurements under low humidity conditions.

(2) Capacitive metal oxide sensors: The operating principle of capacitive metal oxide relative humidity sensors is similar to that of capacitive polymer relative humidity sensors. It mainly utilizes the high thermal expansion and contraction properties of metals. As a result, the measured dew point values can be as low as (-100°C), but this type of sensor is not suitable for high humidity environments.

(3) Capacitive polymer sensors: Capacitive polymer relative humidity sensors use a polymer thin film combined with electrodes and placed between upper and lower electrodes. When water vapor is absorbed into the thin film, changes in relative humidity alter ionic conductivity and change the dielectric constant. As relative humidity increases, capacitance increases accordingly. Like other types of relative humidity sensors, exposure to contamination and extreme environmental conditions may reduce accuracy. However, due to improvements in materials, capacitive relative humidity sensors can operate at high temperatures (100℃) and are less affected by thermal expansion and contraction.

To effectively determine whether temperature and humidity sensors meet acceptable error ranges within their sensing humidity range, this system is designed with humidity control capability, a real-time monitoring human–machine interface, and automatic export of inspection reports for automated temperature and humidity quality inspection software and hardware. Through a current-to-voltage converter and data acquisition module, the temperature and humidity values detected by the sensor are displayed on the human–machine interface. At system startup, relevant parameters must be set. The system then enters the automated quality inspection process. After stabilization time and completion of all measurement points, the system automatically exports inspection reports and generates data curves to provide reference information for quality control personnel.

2. Characteristics of Moist Air Parameters Humidity in meteorology generally refers to atmospheric humidity, representing the amount of water vapor in the air. It is related to the mass of water vapor in the air and is also associated with the ratio between vapor pressure and saturated vapor pressure. Air containing extremely low or nearly zero water vapor is referred to as dry air and can generally be considered an ideal gas. Once the water vapor content increases, its ideal gas characteristics decrease.

In the ASHRAE psychrometric chart shown in Figure 1 [4], the vertical axis represents humidity ratio, and constant humidity ratio lines appear as horizontal lines on the chart. The horizontal axis represents dry-bulb temperature, and constant dry-bulb temperature lines appear as straight rising lines.

Simplified ASHRAE Psychrometric Chart