2013/10 Machinery Monthly - Automatic Quality Inspection System for Capacitive Polymer Dew Point Meters
Development of an Automatic Quality Inspection System for Capacitive Polymer Dew Point Meters (Self-authored)
Jui-Ching Chiu¹, Chen-Ting Wu¹, Ming-Chung Tsai¹, Jen-Cheng Chu²
¹ Graduate Institute of Automation and Control, National Taiwan University of Science and Technology
² Yitian Technology Co., Ltd.
Abstract
The objective of this study is to develop an automatic quality inspection system for capacitive polymer dew point meters under controlled low-humidity environmental conditions. The system consists of a PC controller (with a human–machine interface), an experimental chamber, a dual-flow valve control module, a dew point sensor, a dew point hygrometer (MBW 973), a dew point meter microprocessor chip module, a data acquisition module, a communication system, and a linear compensation system.
In this study, an experimental chamber capable of controlling the dew point measurement range was first constructed. The dew point sensor was installed inside the chamber to perform dew point range inspection. Voltage signals from the PC are output through a DAQ card to control dual-flow valves and regulate the chamber humidity within a range of −60 °C to 20 °C. A chilled-mirror dew point hygrometer (MBW 973) is used as the reference standard.
Servo feedback control is applied to regulate the flow control valves of dry and wet air in order to achieve mixed-flow measurements ranging from low dew points to high dew points. Based on the IEC 61298-2 standard, the system performs automatic quality inspection by controlling output signals under different humidity conditions.
The system allows users to configure parameters through the human–machine interface, including inspection parameters, quality inspection modes, and reference source parameters. It also automatically records data and generates quality control reports to complete the inspection process.
Through automated range-testing experiments, this study establishes the output matching and feedback mechanism of the dry and wet control valves, and determines the flow control parameters of the dew point testing chamber. The target dew point for each measurement range can be maintained within ±1 °C through servo feedback control.
Keywords: humidity, dew point, capacitive polymer dew point meter, data acquisition, quality inspection.
1. Introduction
Humidity is an important environmental physical parameter in many industries, such as the paper industry, food processing industry, and electronics industry. If the generation of water vapor is neglected, it may cause damage to instruments or equipment, or corrosion in pipelines. Therefore, humidity measurement and monitoring have become increasingly important in various industries. In recent years, research topics and measurement technologies related to temperature, humidity, and dew point have become increasingly popular [1–8].
Currently, the most common electronic relative humidity sensors are resistive and capacitive types. These two types are classified according to the changes in electrical properties of their sensing elements. Sensors that use resistance variation as the signal are referred to as resistive sensors, while those that use capacitance variation as the signal are referred to as capacitive sensors. Different fabrication methods result in different characteristics, each having its own advantages and disadvantages depending on the user’s requirements [6,9].
Resistive relative humidity sensors typically use polymers as substrates combined with electrodes. When relative humidity changes, ionic conductivity varies and results in resistance changes. These sensors can produce stable signal variations based on stable polymer materials. However, highly corrosive environments may still damage the sensing elements.
Generally, resistive relative humidity sensors perform poorly in environments below 15% RH, although most relative humidity sensors show similar limitations. Near 100% RH, resistive relative humidity sensors exhibit good performance [4], which is not easily matched by capacitive relative humidity sensors.
Capacitive polymer sensors are commonly used for relative humidity detection, and dew point detection can be achieved through simple temperature measurement on the device under test. Capacitive polymer relative humidity sensors consist of a polymer thin film combined with electrodes and placed between upper and lower electrodes. When water vapor is absorbed by the thin film, changes in relative humidity alter ionic conductivity and modify the dielectric constant.
In general, materials used in such humidity sensors must possess a low dielectric constant (approximately 3–4 Fm⁻¹) and high electrical resistance. Since the dielectric constant of water is approximately 80 Fm⁻¹, the dielectric constant increases as the material absorbs more moisture. Consequently, capacitance variation has a proportional linear relationship with environmental relative humidity [9].
2. Capacitive Polymer Dew Point Meters and Stable Humidity Environment Control
2.1 Capacitive Polymer Relative Humidity Sensors and Dew Point Meters
The basic structure of a capacitive polymer relative humidity sensor consists of a polymer thin film combined with electrodes placed between upper and lower electrodes and fabricated on glass or ceramic substrates, as shown in Figure 1 [10].
When water vapor is absorbed by the thin film, changes in relative humidity alter ionic conductivity and modify the dielectric constant. As relative humidity increases, the capacitance value also increases, as shown in Figure 2 [11].
This type of humidity sensor, similar to other relative humidity sensors, may experience reduced accuracy when exposed to contamination or extreme environmental conditions. However, due to improvements in materials, capacitive relative humidity sensors can operate in high-temperature environments (up to 100 °C) and provide excellent linearity, which is one of their advantages.
There are several methods for detecting capacitance changes. For example, an inductance–capacitance (L–C) oscillation circuit can be used as the primary structure, where humidity is detected by observing changes in oscillation frequency. Another method measures humidity based on variations in frequency [11].
Download for more detail