2014/01 Machinery Monthly - Automatic quality inspection system for capacitive polymer dew point meter

Development of an Automatic Quality Inspection System for Capacitive Polymer Dew Point Meters

 

By Chiu Jui-Ching, Wu Chen-Ting, Tsai Ming-Chung, and Chu Jen-Cheng


The purpose of this study is to construct a low-humidity environment control and an automatic quality inspection system for capacitive polymer dew point meters. The system includes a PC controller (with human–machine interface), experimental chamber, dual-flow valve control module, dew point sensor, dew point hygrometer MBW 973, dew point meter microprocessor chip module, data acquisition module, communication system, and linear compensation system. In this study, an experimental chamber capable of controlling the dew point measurement range was first constructed. The dew point sensor was installed in the experimental chamber to perform dew point range testing. The PC outputs voltage signals through a DAQ card to control the dual-flow valves to regulate the chamber humidity (-60℃~20℃), while the chilled mirror dew point hygrometer MBW 973 is used as the reference standard. Servo feedback control is applied to regulate the dry and wet air flow control valves to achieve mixed-flow measurements from low dew point to high dew point. According to the IEC 61298-2 standard, the system performs automatic quality inspection by controlling output signals under different humidity conditions. The system allows parameter settings through the human–machine interface including inspection parameters, quality inspection modes, and standard reference parameters, and automatically records data and outputs quality control reports to complete the inspection process. Through automated range testing experiments, this study establishes the matching output and feedback mechanism of the dry and wet control valves, and determines the flow control parameters of the dew point testing chamber. The target dew point of each measurement range can be controlled within ±1℃ through servo feedback control.

 

Keywords: humidity, dew point, capacitive polymer dew point meter, data acquisition, quality inspection.


Introduction

Humidity is an important environmental physical parameter in many industries, such as the paper industry, food industry, and electronics industry. If the generation of water vapor is ignored, it may cause damage to instruments or equipment, or corrosion of pipelines. Therefore, humidity measurement and monitoring have become increasingly important in various industries. In recent years, research topics and measurement technology developments related to temperature, humidity, and dew point have gradually increased in popularity [1-8].
Currently, the most common electronic relative humidity sensors are resistive and capacitive types. These two types of humidity sensors are classified according to the changes in electrical properties of the sensing elements. If the signal is based on resistance variation it is classified as resistive, and if the signal is based on capacitance variation it is classified as capacitive. Different fabrication methods result in different characteristics, each having its own advantages and disadvantages depending on the user's requirements [6,9].
Resistive relative humidity sensors use polymers as substrates combined with electrodes. When relative humidity changes, ionic conductivity changes, resulting in resistance variation. Resistive elements can produce stable signal variations based on stable polymer materials; however, highly corrosive environments can still damage the sensing elements. Generally, resistive relative humidity sensors perform poorly in environments below 15% RH, though most relative humidity sensors have similar limitations. Near 100% RH, resistive relative humidity sensors perform well [4], which is not comparable to capacitive relative humidity sensors.
Capacitive polymer sensors are commonly used for relative humidity detection, and by using simple temperature measurement on the device under test, dew point detection can also be achieved. Capacitive polymer relative humidity sensors use a polymer thin film combined with electrodes and placed between upper and lower electrodes. When water vapor is absorbed into the thin film, changes in relative humidity alter ionic conductivity and change the dielectric constant. Typically, the materials used for such humidity sensors require a low dielectric constant (approximately 3~4 Fm-1) and high resistance. Since the dielectric constant of water is 80 (Fm-1), when the material absorbs moisture, the dielectric constant increases with environmental humidity. This phenomenon creates a proportional linear relationship between capacitance variation and relative humidity [9].

 

Capacitive Polymer Dew Point Meters and Stable Humidity Environment Control

Capacitive Polymer Relative Humidity Sensors and Dew Point Meters

The basic structure of a capacitive polymer relative humidity sensor uses a polymer thin film combined with electrodes and placed between upper and lower electrodes, fabricated on glass or ceramic substrates as shown in Figure 1 [10]. When water vapor is absorbed into the thin film, changes in relative humidity alter ionic conductivity and change the dielectric constant. As relative humidity increases, the capacitance value also increases, as shown in Figure 2 [11]. This type of humidity sensor, like other relative humidity sensors, may experience reduced accuracy when exposed to contamination or extreme environmental conditions. However, due to improvements in materials, capacitive relative humidity sensors can operate in high-temperature environments (100℃) and offer good linearity, which is one of their advantages.
There are several methods to detect capacitance changes. For example, using an inductance-capacitance (L-C) oscillation circuit as the main structure, humidity can be detected through changes in oscillation frequency. Another method measures humidity based on frequency variation [11]. Capacitive relative humidity sensors use a polymer thin film placed between electrodes. When water vapor is absorbed by the thin film, the dielectric constant changes, resulting in capacitance variation. The calculation formula for the capacitive relative humidity sensor is as follows [6,9]:

Figure 1. Structure of a Capacitive Polymer Humidity SensorFigure 1. Structure of a Capacitive Polymer Humidity Sensor

Download Link :