Gas flow standard calibration system

Gas Flow Calibration System | ISO 9300 Critical Flow Venturi Nozzle High Accuracy Calibration

This Gas Flow Calibration System is designed in accordance with ISO 9300 standards, utilizing critical flow Venturi nozzles and sonic nozzles to provide high-accuracy gas flow calibration solutions. It ensures excellent measurement accuracy, stability, and traceability, making it ideal for semiconductor, laboratory, and industrial process applications.

Gas flow calibration system design ISO 9300 critical flow Venturi nozzle
Gas Flow Calibration System Design

ISO 9300 Critical Flow Venturi Nozzle Measurement Standard

This system is developed in compliance with ISO 9300 (Critical Flow Venturi Nozzle Flow Measurement), configuring multiple sonic nozzles across different flow ranges to establish standardized calibration points. It ensures high accuracy and consistency in gas flow calibration aligned with international standards.

System Specifications and Flow Range

  • Flow Range: 1 ~ 1000 m³/h (4 nozzle sets: 40 / 160 / 320 / 480 m³/h)
  • Pipe Size: DN15 ~ DN150 (DN150 ≈ 150 mm ≈ 6 inch)
  • Maximum Flow:
    • 60 Hz + heater: 539.3 Nm³/h
    • Without heater: up to 1000 Nm³/h (standard condition at 20°C)
  • Control Interface: 4–20 mA analog input
  • Measurement Uncertainty: ≤ 0.2%
  • Piping Design: Vertical parallel configuration

Principle of Critical Flow Venturi and Sonic Nozzle Measurement

This gas flow calibration system uses multiple critical flow Venturi nozzles to generate a stable reference flow. When the gas reaches the critical pressure ratio at the nozzle throat, the flow velocity becomes sonic (critical flow condition). Under this condition, the flow rate depends only on upstream pressure and temperature and is independent of downstream conditions.

Based on the principle of mass conservation:
Flow through the device under test = Flow through the sonic nozzle

The system integrates temperature, pressure, and humidity sensors. Data is collected and compensated through a computer system to calculate standard flow values and achieve high-accuracy gas flow calibration.

Advantages of Sonic Nozzle Calibration Technology

  • Independent of downstream pressure: Ensures stable and high-accuracy measurement
  • ISO 9300 compliant: Provides international traceability
  • No moving parts: Long-term stability with minimal maintenance
  • Excellent repeatability: Ideal for precision calibration applications

Speed of Sound and Flow Calculation

Speed of sound formula:
c = 331.6 + 0.6T (m/s) (T in °C)

At 20°C, the speed of sound is approximately 343.6 m/s.

Flow calculation principle:
Flow Rate = Cross-sectional Area × Flow Velocity

Applications

  • Semiconductor process gas calibration
  • Laboratory flow calibration and testing
  • Industrial process gas monitoring
  • Precision instrument calibration

Why Choose Our Gas Flow Calibration System

  • Compliant with ISO 9300 international standards
  • Sonic nozzle technology ensures high stability and accuracy
  • Suitable for semiconductor and advanced industrial applications

Download more information