Huba 630 differential pressure switch

Medium : Liquids and neutral gases
Druckbereich : 6 ... 5500 mbar
Pressure range : Screw terminals, Tab connectors 6.3 mm
Pressure connection : Inside thread

Features

The differential pressure switches of Type 630 are suitable for flow monitoring in sanitary and heating systems or for level control in general process engineering. The extremely robust design allows for a system pressure and a one-sided overload of up to 20 bar.

• High overpressure safety on both connections (P1, P2) up to 10/20 bar
• Functionally simple and robust mechanics with high operational reliability
• Also suitable for slightly aggressive liquids and gases
• Reproducibility up to < ±0.4 mbar

Specification

Pressure range
Relative / Differential6 ... 5500 mbar
Operating conditions
MediumLiquids and neutral gases
Medium temperatureNBR 0 ... +80°C
FPM -10 ... +80°C
EPDM -10 ... +80°C
Silicone -40 ... +80°C
Ambient temperature+65°C
Storage temperature-40 ... +80°C
Tolerable overload (P1 > P2)< 200 mbar: 10 bar
> 200 mbar: 20 bar
Rupture pressure30 bar
Lowest turn-on pressure> 6 mbar
Smallest switching difference> 3 mbar
Materials in contact with medium
DiaphragmNBR / EPDM / FPM / Silicone
Pressure caseAnodized aluminium / Brass / Nickel plated brass
Other componentsX14CrMoS17 (1.4104)
X5CrNi18-10 (1.4301)
X10CrNi18-8 (1.4310)
Steel A2 screws
Polyacetate-C / Polyamide
Electrical overview
Nominal voltage250 VAC
Nominal current (resistive)1 A
Nominal current (motor)0.5 A
Contact systemChangeover contact
Mechanical life10⁶ switching cycles
Protection
Without coverIP00
With cover (PG11)IP54
With cover (PG9)IP65
Repeatability
NBR / Silicone±5% of switching point (min ±0.4 mbar)
FPM / EPDM±10% of switching point (min ±0.8 mbar)
Electrical connections
Screw terminalsOption
Tab connectorsAMP 6.3 mm
Cable glandPG9 / PG11
Pressure connections
ThreadG⅛
Straight screwed connectionØ 6 mm (G⅛)
Screwed socketØ 6 mm (G⅛)
Weight
Aluminium base~ 380 g
Brass / Nickel plated brass~ 1000 g

Download

Note : Design and specifications are subject to change without notice.
Note : Please make sure the product and the device which connect with RS-485 are on common ground, avoid damaged product.