Huba 528 OEM type Pressure Transmitter

Huba 528
  • Huba 528 OEM type Pressure Transmitter
Features
The compact type 528 pressure transmitter is based upon the well proven ceramic technology developed by Huba Control over 20 years ago. These transmitters are suitable for applications across a broad spectrum of industries.

• Compact, rugged construction
• Negligible temperature influence on accuracy
• Large selection of connections available
• Saving time by quick cable mounting by the customer with swift connector
Applications
Solar heating / CDU / In-Rack CDU / Circulation cooler / Large heat pumps / Water pumps / Water treatment / Mobile hydraulics / Industrial automation
Specification
Product Features
Pressure range
Relative: -1 ... 0 ... 60 bar
Absolute: 0 ... 1 ... 16 bar
Accuracy
Characteristic line a: ± 0.3% fs
Resolution: 0.1% fs
Thermal characteristic (max.): ± 0.2% fs / 10K
Long term stability acc. EN 60770-1 (max.): ± 0.25% fs
Operating Conditions
Medium
Liquids and gases
Temperature
Medium:
FPM: -15 ... +125 °C (-15 ... +120 °C)
EPDM: -40 ... +125 °C (-30 ... +120 °C)
NBR: -20 ... +100 °C
MVQ: -40 ... +125 °C (-30 ... +120 °C)
Ambient: -30 ... +85 °C (-25 ... +85 °C)
Storage: -50 ... +100 °C
Tolerable overload / Rupture pressure
< 4 bar: 3.0 x fs
> 4 bar: 2.5 x fs
Materials
Cover
Stainless steel 1.4404 / AISI 316L
Plug accommodation
Polyarylamide 50% GF UL 94 V-0
Materials in contact with medium
Pressure connection: Stainless steel 1.4404 / AISI 316L / PVDF
Sensor: Ceramic Al2O3 (96%)
Sealing material: FPM, EPDM, NBR, MVQ
Electrical Data
Output
2 wire: 4 ... 20 mA (7 ... 33 VDC / 10 ... 30 VDC)
3 wire: 0 ... 5 V; 1 ... 6 V; 0 ... 10 V; ratiom. 10 ... 90%
Load / Current consumption
Voltage outputs: > 10 kOhm / < 100 nF
Current consumption: < 7 mA (voltage output)
< 23 mA (4 ... 20 mA)
Polarity reversal protection
Short circuit proof and protected against polarity reversal.
Each connection protected against crossover up to max. supply voltage.
Insulation voltage
500 VDC
Response time
< 2 ms (1 ms typ.)
Load cycle < 100 Hz
Electrical connection
Swift connector (with or without cable 1.5 / 2.0 / 3.0 / 5.0 m)
Connector DIN EN 175301-803-A or C
Metri Pack 150 P2S Series
Connector M12x1
Braids
Connector RAST 2.5 (3 wire only)
Protection class
IP65 / IP67 depending on connection
Installation
Installation arrangement
Unrestricted
Recommendation: Pressure connection facing downwards
Approvals
EMC
CE conformity acc. EN 61326-2-3
Enhanced EMC protection EN 50121-3-2
Shock
IEC 68-2-27 100 g, 11 ms half sine wave
Vibration
IEC 68-2-6 20 g, 15 ... 2000 Hz
UL
ANSI/UL 61010-1 acc. E325110
Drinking water approval
NSF/ANSI 61 / 372 acc. MH60087
Explosion protection
Ex II 1/2 G Ex ia IIC T4 Ga/Gb
Ex II 1/2 D Ex ia IIIC T125°C Da/Db
ATEX / IECEx
Weight
~ 90 g
Download
Note : Design and specifications are subject to change without notice.
Note : Please make sure the product and the device which connect with RS-485 are on common ground, avoid damaged product.